Product overview
The functions of the kdzd5550 series voltage blow-out tester are described as electrical insulation strength tester, medium strength tester, etc. According to the national gb1408. 1-2006 insulation material electrical intensity test method. The working method is that adding a higher than normal voltage to the insulation of the measured equipment for a specified period of time, and adding to the voltage, would produce only a small leak, which would be better insulated. The three modules of the programme power source module, the signal capture reconciliation module and the computer control system form a test system with alarm and time control features。
The kdzd5550 series of electro-pressure piercing test instruments are applied primarily to the testing of solid insulation materials (e. G. Plastics, rubber, laminated materials, thin membranes, resins, clouds, ceramics, glass, insulation paints, etc., and insulations) under industrial frequency or current voltage. Designed for use in high voltage penetration tests in electrical and electronics industries, the ministry of mechanical industry, research institutes, military engineering research, insulation material production units。
Technical parameters
Pressure level: 5kv, 10kv, 20kv, 50kv, 100kv, 150kv
Electricity: 5kv
Input voltage: ac220v 50hz
Output voltage: ac: 0-5kv; dc:0-5kv
Output power: 0. 5 kva
Measurement range: ac: 0. 05-5kv; dc:0-5kv
Measurement error: 2%
Increment rate: 0. 025 kv/s ~ 0. 5/s
Electricity: 10kv
Input voltage: ac220v 50hz
Output voltage: ac: 0-10 kv; dc: 0-10kv
Output power: 1 kva
Measurement range: ac: 0. 5-10 kv; dc: 0. 5-10kv
Measurement error: 2%
Increment rate: 0. 05 kv/s ~ 1 kv/s
Equipment self-esteem: 150 kg
Electricity: 20kv
Input voltage: ac220v 50hz
Output voltage: ac: 0-20kv dc:0~20kv
Output power: 2 kva
Measurement range: ac: 1-20 kv; dc: 1-20kv
Measurement error: 2%
Increment rate: 0. 1 kv/s ~2kv/s
Equipment self-esteem: 150 kg
Electricity: 50kv
Input voltage: ac220v 50hz
Output voltage: ac: 0 ~ 50 kv; dc: 0 ~ 50 kv
Output power: 3 kva
Measurement range: ac: 1 ~ 50 kv; dc: 1 ~ 50 kv
Measurement error: 2%
Increment rate: 0. 20 kv/s ~ 5kv/s
Equipment self-esteem: 150 kg
Electricity: 100kv
Input voltage: ac220v 50hz
Output voltage: ac: 0 ~ 100 kv; dc: 0 ~ 100kv
Output power: 10 kva
Measurement range: ac1-100kv; dc1-100kv
Measurement error: 2%
Increment rate: 0. 02 kv/s ~ 10 kv/s
Equipment self-esteem: 300kg
Electricity: 150 kv
Input voltage: ac220v 50hz
Output voltage: ac: 0 ~ 150 kv; dc: 0 ~ 150 kv
Output power: 15 kva
Measurement range: ac1-150kv; dc1-150kv
Measurement error: 2%
Increment rate: 1 kv/s ~ 10 kv/s
Equipment self-esteem: 300kg
Pressure tolerance time: 0-4h (aerial)
Leakage: allows computer software to set free
Power: exchange of 220 v (1)(10%) single-phase communication voltage and 50 hz (1)(%) frequency
Test environment: temperature: 15 ~ 30°c, relative humidity: 30% ~ 65% stable operation。
Outer size: long x width x height: 1000 x 1000 x 1400 mm
Grounding requirements: the instrument needs to be located separately to conform to national standards, with metal rods buried below a minimum of 1. 5 metres deep
Product characteristics
Operating control by equipment for touch screens。
2. If a curve analysis is required, it can be equipped with a computer, with only data and curve recording functions, without equipment control, avoiding alternation and humanization of the test personnel between computers and equipment。
3. The equipment has the memory function of test parameters, the same test conditions need not be set for each test and the last test setting parameters will be remembered for power outage。
4. The test interface is simple and clear and is accompanied by an indicative curve indicating that different parameters and curves are moving differently to facilitate understanding。
5. Control panels are simple, functionally stated and simple。
6 ten test records can be recorded and displayed simultaneously to facilitate comparative analysis of test data. And you can discard undesirable data in real time。
7. The equipment test interface is presented in real time and is more accessible to the test process。
The device has a security warning that the test cannot begin without closing the test box door and will eject a warning, eject a warning when full (i. E. High pressure transformer has no output) and if the test is opened, the test will automatically end。
The equipment is equipped with a three-coloured alarm. (selection)
Product packing
The kdzd5550 series voltage piercing tester: 1
Enter output connection line: 1 set
Equipment surface line: 1
Test electrodes: 2
Product description: 1
Product conformity certificate: 1